FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 2081570
·
Received April 13, 2011
Report
- Report Number
- 9615739-2011-00092
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- March 21, 2011
- Report Date
- March 21, 2011
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE SIDECOM BOARD WAS REPLACED TO REPAIR THE BED.
Description of Event or Problem · 1
THE ACCOUNT STATED THAT THE PATIENT POSITIONING MONITOR (PPM) WAS ARMED AND ALARMED IN THE ROOM BUT NO NURSE CALL WAS PLACED TO THE NURSE STATION. NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAREASSIST BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |