FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 2081570 · Received April 13, 2011

Report

Report Number
9615739-2011-00092
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
March 21, 2011
Report Date
March 21, 2011
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SIDECOM BOARD WAS REPLACED TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT THE PATIENT POSITIONING MONITOR (PPM) WAS ARMED AND ALARMED IN THE ROOM BUT NO NURSE CALL WAS PLACED TO THE NURSE STATION. NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1 UNK