STEALTHSTATION S7 SYSTEM
MDR report key: 2081510
·
Received April 19, 2011
Report
- Report Number
- 1723170-2011-00891
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 24, 2011
- Report Date
- March 24, 2011
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
SOFTWARE INVESTIGATION WAS COMPLETED. ISSUE DETERMINED TO BE DUE TO VIEWS LAYOUT IN TRANSLATED LANGUAGES. THIS ISSUE WILL BE CONTINUALLY MONITORED IN A MEDTRONIC SOFTWARE ANOMALY TRACKING DATABASE.
Description of Event or Problem · 1
A MEDTRONIC REPRESENTATIVE REPORTED THAT HE INSTALLED SYNERGY CRANIAL 2.1.5 WITH STEALTHMERGE AND STEALTH 3D IN A ROLLING STONE COMPUTER SYSTEM. IN THE PLAN VIEW, WHEN THE MEDTRONIC REPRESENTATIVE SELECTED THE 6 VIEWS BOX AND TRIED TO ADD THE VIEW TYPES, THE SOFTWARE FROZE AND HE HAD TO RE-START BY PRESSING CTRL + ALT + BACKSPACE. THE MEDTRONIC REP STATED THAT THE BEHAVIOR IS THE SAME SELECTING OTHER VIEWS OR IN NAVIGATE. THERE WAS NO PT PRESENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEALTHSTATION S7 SYSTEM | STEREOTAXIC INSTRUMENT (HAW) | HAW | MEDTRONIC NAVIGATION, INC. | S7 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NONE |