FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TRIA NAVIGATION SYSTEM
MDR report key: 2081196
·
Received April 21, 2011
Report
- Report Number
- 1723170-2011-00809
- Event Type
- Malfunction
- Date Received
- April 21, 2011
- Date of Event
- January 14, 2010
- Report Date
- January 14, 2010
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K001153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
AS RECEIVED, THE PSU HAD SEVERAL BROKEN STANDOFFS INSIDE. DURING A FUNCTIONAL TEST, THE POSITION SENSOR UNIT (PSU) RETURNED HIGH GEOMETRY FOR AN ACTIVE FRAME. THE VOLUME WAS ONLY SLIGHTLY REDUCED AT THE BACK EDGE FOR THE PASSIVE PROBE. THERE WERE TOO FEW MARKERS VISIBLE TO RUN THE AAK TEST. THE PSU IS OUT OF CALIBRATION.
Description of Event or Problem · 1
A MEDTRONIC REP REPORTED, THE CAMERA ON THE SITE'S TRIA SYSTEM WAS NOT TRACKING THE INSTRUMENTATION. THIS EVENT DID NOT OCCUR DURING SURGERY AND NO PT WAS PRESENT.