FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION TRIA NAVIGATION SYSTEM

MDR report key: 2081196 · Received April 21, 2011

Report

Report Number
1723170-2011-00809
Event Type
Malfunction
Date Received
April 21, 2011
Date of Event
January 14, 2010
Report Date
January 14, 2010
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
PMA / PMN Number
K001153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AS RECEIVED, THE PSU HAD SEVERAL BROKEN STANDOFFS INSIDE. DURING A FUNCTIONAL TEST, THE POSITION SENSOR UNIT (PSU) RETURNED HIGH GEOMETRY FOR AN ACTIVE FRAME. THE VOLUME WAS ONLY SLIGHTLY REDUCED AT THE BACK EDGE FOR THE PASSIVE PROBE. THERE WERE TOO FEW MARKERS VISIBLE TO RUN THE AAK TEST. THE PSU IS OUT OF CALIBRATION.

Description of Event or Problem · 1

A MEDTRONIC REP REPORTED, THE CAMERA ON THE SITE'S TRIA SYSTEM WAS NOT TRACKING THE INSTRUMENTATION. THIS EVENT DID NOT OCCUR DURING SURGERY AND NO PT WAS PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEALTHSTATION TRIA NAVIGATION SYSTEM STEREOTAXIC INSTRUMENT (HAW) HAW MEDTRONIC NAVIGATION, INC. TRIA NA