FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2081034 · Received April 14, 2011

Report

Report Number
9680959-2011-01068
Event Type
Malfunction
Date Received
April 14, 2011
Date of Event
March 25, 2011
Report Date
April 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. TWO FUSES WERE REPLACED AND A REQUEST WAS MADE TO HAVE AN ELECTRICIAN EVALUATE THE POWER USAGE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE 7900 SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1