FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2081028 · Received April 14, 2011

Report

Report Number
1720753-2011-03600
Event Type
Malfunction
Date Received
April 14, 2011
Date of Event
March 25, 2011
Report Date
April 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MFR'S SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REP REPLACED THE UNINTERRUPTED POWER SUPPLY AND POWER CORD. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM WOULD SHUT OFF BY ITSELF WHEN POSITIONING FOR PROCEDURE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1