FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2081010 · Received April 14, 2011

Report

Report Number
1720753-2011-03592
Event Type
Malfunction
Date Received
April 14, 2011
Date of Event
January 10, 2011
Report Date
April 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY CABLE AND THE INTERCONNECT CABLE WERE REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE INTERCONNECT CABLE FAILED ON THE 9600 SYSTEM. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1