FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2081008 · Received April 14, 2011

Report

Report Number
9680959-2011-01069
Event Type
Malfunction
Date Received
April 14, 2011
Date of Event
March 30, 2011
Report Date
April 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CABLE FROM THE MONITOR CART TO THE C-ARM HAD DAMAGED PINS AND WAS REPLACED. THE FUSE IN THE MONITOR CART WAS REPLACED AND THE INTERCONNECT CABLE FROM THE C-ARM TO THE MONITOR NEEDED TO BE ORDERED. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THERE WAS NO POWER TO THE MONITOR CART. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1