FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2080999 · Received April 14, 2011

Report

Report Number
1720753-2011-03581
Event Type
Malfunction
Date Received
April 14, 2011
Date of Event
March 28, 2011
Report Date
April 14, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE CAMERA AND ADJUSTED THE TRACKING. THE SYSTEM WAS TESTED AND IS OPERATING AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE MILLIAMP VOLTAGE ON THE 9800 SYSTEM WAS TOO HIGH, AND THE IMAGES ON THE MONITOR APPEARED WHITE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1