FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2080910
·
Received May 7, 2011
Report
- Report Number
- 2050012-2011-01500
- Event Type
- Malfunction
- Date Received
- May 7, 2011
- Date of Event
- March 21, 2011
- Report Date
- April 6, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RESULT: ERRONEOUS DATA GENERATED.CONCLUSION: SAMPLE INTEGRITY AND/OR MAINTENANCE ISSUES MAY HAVE BEEN CONTRIBUTING FACTORS TO THE EVENT.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT 14 ERRONEOUSLY LOW OR HIGH GLUCOSE RESULTS WHEN RUN IN DUPLICATE MODE ON A UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE CUSTOMER NORMALLY RUNS PATIENT SAMPLES IN DUPLICATE MODE AND VERIFIES THE RESULTS PRIOR TO REPORTING. THE SUSPECT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. PATIENT TREATMENT WAS NOT IMPACTED. THE CUSTOMER HAS EXPERIENCED SAMPLE INTEGRITY ISSUES WHICH MAY HAVE CONTRIBUTED TO THE ERRONEOUSLY LOW AND HIGH GLUCOSE RESULTS OBSERVED. A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |