UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.
Report
- Report Number
- 2050012-2011-01415
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 5, 2011
- Report Date
- April 5, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
CUSTOMER TECHNICAL SUPPORT (CTS) INFORMED CUSTOMER THAT THE ABOVE LEAK CAN BE DUE TO A CLOGGED DRAIN TUBE, OR, DUE TO A FAULTY DRAIN VALVE. CTS OFFERED TO CALL THE FIELD SERVICE; HOWEVER, THE CUSTOMER DECLINED. CTS REFERRED THE CUSTOMER TO THE INSTRUCTIONS IN THE IFU TO CLEAN THE ISE DRAIN. CTS STATED THE ISE DRAIN IS STILL OVERFLOWING POST CLEANING. CUSTOMER DISCONNECTED THE DRAIN TUBING FROM THE SOLENOID AND IT STARTED TO DRAIN. CUSTOMER STATED THAT THEIR OWN BIOMED FOUND A CLOGGED LINE. NO SPECIFIC INFORMATION PROVIDED. NO FURTHER LEAKING HAS BEEN NOTED AS OF (B)(4) 2011. CUSTOMER DECLINE SERVICE
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ION-SELECTIVE ELECTRODE (ISE) DRAIN IS OVERFLOW. NO INJURY OR EXPOSURE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC. | DXC800 PRO | N/A |