UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2080873 · Received May 6, 2011

Report

Report Number
2050012-2011-01498
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
March 18, 2011
Report Date
April 6, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULT: ERRONEOUS DATA GENERATED. CONCLUSION: SAMPLE INTEGRITY AND/OR MAINTENANCE ISSUES MAY HAVE BEEN CONTRIBUTING FACTORS TO THE EVENT.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT ONE ERRONEOUSLY HIGH GLUCOSE RESULT WHEN RUN IN DUPLICATE MODE ON A UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE INITIAL GLUCOSE RESULT YIELDED 80 MG/DL. A REPEAT ANALYSIS FOR GLUCOSE YIELDED 71 MG/DL. AN ADDITIONAL REPEAT ANALYSIS WAS PERFORMED ON THE SAME ANALYZER AND YIELDED 75 MG/DL WHICH WAS REPORTED OUT OF THE LABORATORY (THE OTHER TWO PREVIOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY). PATIENT TREATMENT WAS NOT IMPACTED. THE CUSTOMER HAS EXPERIENCED SAMPLE INTEGRITY ISSUES WHICH MAY HAVE CONTRIBUTED TO THE ERRONEOUSLY HIGH GLUCOSE RESULT OBSERVED. A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED.