FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2080845 · Received May 6, 2011

Report

Report Number
2531779-2011-03211
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 9, 2011
Report Date
April 9, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(6) 2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS:A REVIEW OF PUMP HISTORY REVEALED A LOW BATTERY WARNING DUE TO LOW VOLTAGE. THE PUMP POWERED ON NORMALLY AND WAS EXERCISED FOR 5 HOURS WITHOUT ANY OVERHEATING OR ALARMS. THE PUMP ELECTRICAL CURRENT DRAW WAS TESTED AND CONFIRMED TO BE LESS THAN 1 A. THE PUMP WAS OPENED NO DAMAGE OR DEFECTS WERE FOUND TO THE POWER FLEX, BATTERY CONNECTIONS AND INTERNAL COMPONENTS.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILLED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP HAD A SHORT BATTERY LIFE ISSUE (LASTED ONLY 2 DAYS) AND THE BATTERY FELT WARM. THE PUMP REPORTEDLY HAS NO PHYSICAL DAMAGE AND THE BATTERY CAP WAS SECURELY ATTACHED. THE REPORTED ISSUES WERE NOT RESOLVED WITH TROUBLESHOOTING WITH ANIMAS CUSTOMER SUPPORT. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. A REPLACEMENT PUMP WAS SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR