AU681-02E CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-01516
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 6, 2011
- Report Date
- April 6, 2011
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K961274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE OPERATOR LOADED A FRESH CREATININE REAGENT CARTRIDGE WITHOUT CALIBRATING AND RUNNING CONTROLS AS REQUIRED IN THE INSTRUCTIONS FOR USE. LABORATORY PRACTICE CONTRIBUTED TO ERROR BY FIXING REAGENT ON BOARD INSTRUMENT TO DISABLE ON BOARD LIMITS TO REAGENT USE AND THE AUTOMATIC REQUIREMENT FOR CALIBRATION OF NEW BOTTLES OF CREATININE REAGENT. USER ERROR IS THE ROOT CAUSE OF THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) IN REGARDS TO OBTAINING AN ERRONEOUSLY HIGH SERUM CREATININE (CRE) RESULT AND AN ERRONEOUSLY LOW CALCULATED ESTIMATED GLOMERULAR FILTRATION RATE (EGFR) RESULT THAT WERE GENERATED BY THE AU681-02E CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE PATIENT SAMPLE WAS RETESTED AFTER PROPER CALIBRATION AND AMENDED REPORTS WERE RELEASED WITHIN 20 HOURS. ORDERING DOCTORS WERE CONTACTED AND REVISED RESULTS WERE TRANSMITTED. IT IS UNKNOWN IF PATIENT TREATMENT WAS AFFECTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU681-02E CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |