AU681-02E CLINICAL CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-01459
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 6, 2011
- Report Date
- April 6, 2011
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K961274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE OPERATOR LOADED A FRESH CREATININE REAGENT CARTRIDGE WITHOUT CALIBRATING AND RUNNING CONTROLS AS REQUIRED IN THE INSTRUCTIONS FOR USE. LABORATORY PRACTICE CONTRIBUTED TO ERROR BY FIXING REAGENT ON BOARD INSTRUMENT TO DISABLE ON BOARD LIMITS TO REAGENT USE AND THE AUTOMATIC REQUIREMENT FOR CALIBRATION OF NEW BOTTLES OF CREATININE REAGENT. USER ERROR IS THE ROOT CAUSE OF THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI), IN REGARDS TO OBTAINING ONE (1) ERRONEOUSLY HIGH SERUM CREATININE (CRE) RESULT AND ONE (1) ERRONEOUSLY LOW CALCULATED ESTIMATED GLOMERULAR FILTRATION RATE (EGFR) RESULT THAT WERE GENERATED BY THE (B)(4)CHEMISTRY ANALYZER. THE ERRONEOUS RESULT WAS REPORTED OUT OF THE LABORATORY. THE PATIENT SAMPLE WAS RETESTED AND AMENDED REPORTS WERE RELEASED FOR CRE AND EGFR RESULTS. IT IS UNKNOWN IF PATIENT TREATMENT WAS AFFECTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU681-02E CLINICAL CHEMISTRY ANALYZER | AU681 CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | AU681-02E | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |