FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER.
MDR report key: 2080769
·
Received May 6, 2011
Report
- Report Number
- 2050012-2011-01419
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 6, 2011
- Report Date
- April 6, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) INSPECTED WASH/VACUUM PROBES AND FOUND SMALL OBSTRUCTION IN 2ND PROBE FROM RIGHT. FSE REPLACED PROBE AND PERFORMED WASH TOWER ALIGNMENT. FSE FLASHED ALL CHANNELS ON PHOTOMETER, PERFORMED AUTO SET AND SAVE. PERFORMED NECESSARY ALIGNMENTS. THIS ISSUE HAS BEEN RESOLVED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT THE SYSTEM PRODUCED MULTIPLE CUVETTE WATER BLANK FAILURES. THE CUVETTES WERE WET ON THE OUTSIDE AND LIQUID WAS NOTED IN REACTION CAROUSEL COMPARTMENT. NO INJURY OR EXPOSURE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC. | DXC800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |