FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2080742 · Received May 6, 2011

Report

Report Number
2024168-2011-03266
Event Type
Injury
Date Received
May 6, 2011
Date of Event
March 14, 2011
Report Date
April 12, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE NOTED THAT THE PLUNGER, ANTERIOR NEEDLE, LINK AND CUFFS WERE NOT RETURNED WITH THE DEVICE; THEREFORE, THE SCOPE OF THIS INVESTIGATION WAS VERY LIMITED. INSPECTION OF THE RETURNED DEVICE INDICATED THAT THE POSTERIOR CUFF MISS HAD OCCURRED AS EVIDENCED BY DAMAGED POSTERIOR NEEDLE TIP. THERE WAS NO NEEDLE STRIKE MARK OBSERVED AT THE POSTERIOR FOOT. THIS IS INDICATIVE OF THE POSTERIOR NEEDLE BEING DEFLECTED AWAY FROM POSTERIOR FOOT AND INTERACTED WITH HUMAN TISSUE INSTEAD OF ENGAGED WITH THE POSTERIOR CUFF INSIDE THE POSTERIOR FOOT POCKET DURING NEEDLE DEPLOYMENT AS DESIGNED. DURING TESTING, THE PROXY PLUNGER WAS INSERTED TO TEST THE NEEDLE TRAJECTORY AND PUSH MANDREL TRAVEL AND THE RESULTS MET THE MANUFACTURING CRITERIA. BASED ON THE SUCCESSFUL TEST OF THE DEVICES, THE PROBABLE CAUSE FOR THE POSTERIOR CUFF MISS IS NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT AS SPECIFIED IN THE INSTRUCTIONS FOR USE (IFU). NO MANUFACTURING OR QUALITY ISSUE WAS DETECTED. A REVIEW OF THE PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. OVERALL, THERE DOES NOT APPEAR TO BE ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4): DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATED DATE (REPORTED AS OCCURRING 2-3 WEEKS AGO). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP WILL BE SUBMITTED WITH ALL ADDITIONAL, RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN UNKNOWN IF TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, THE PRODUCT EXPERIENCE THAT OCCURRED IS UNKNOWN. THE METHOD OF ACHIEVING HEMOSTASIS IS UNSPECIFIED. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, ADDITIONAL INFORMATION WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 020226H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention