Report Number
2939301-2011-03752
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 14, 2011
Report Date
April 26, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K082590.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(4) 2011)-DEVICE EVALUATION: THE LAY USER/PATIENT'S PRODUCT(S) INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A CORRODED BATTERY CONTACT. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE BATTERY CONTACT ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.