FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2080722 · Received May 6, 2011

Report

Report Number
2531779-2011-03204
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 6, 2011
Report Date
April 6, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. THE KEYPAD BUTTON WAS RESPONDING INTERMITTENTLY. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE CONTACTS.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PATIENT CONTACTED AN ANIMAS REPRESENTATIVE AND ALLEGED THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THAT THE OK BUTTON REQUIRED SEVERAL PRESSES FOR THE PUMP TO RESPOND. HE DENIED THAT THE DEVICE WAS EXPOSED TO WATER OR THAT THE KEYPAD WAS DAMAGED. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS NOT RESPONDING APPROPRIATELY TO KEYPAD BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER, THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 10