FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER.

MDR report key: 2080712 · Received May 6, 2011

Report

Report Number
2050012-2011-01398
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 6, 2011
Report Date
April 6, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE INFORMATION WAS NOT PROVIDED. THE SYSTEM WAS CALIBRATED AT 0215 AND AGAIN AT 0230. PATIENT SAMPLES WERE RUN AND REPORTED. THE SYSTEM WAS RECALIBRATED AT 0808, AND QC RUN AT 0828. THE LAB CONFIRMED THAT THE OPERATOR DID NOT RUN QC AFTER LOADING REAGENT AND RECALIBRATING. THE SAMPLES WERE RERUN AFTER THE SYSTEM WAS RECALIBRATED AT 0808, AND RECOVERY WAS HIGHER, MATCHING THE OTHER INSTRUMENT. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED: WORN O-RINGS. BROKEN NA ELECTRODE CONNECTOR. MODULAR CHEMISTRY SAMPLE PROBE. FSE VERIFIED PERFORMANCE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS LOW SODIUM (NS), POTASSIUM (K), AND CHLORIDE (CL) RESULTS GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. THE LABORATORY REPEATED THE SAMPLES ON AN ALTERNATE UNIT ONCE THEY NOTICED A TREND OF LOW RESULTS. HIGHER RESULTS WERE OBTAINED AND THE AMENDED REPORTS WERE INITIATED. TREATMENT WAS NOT INITIATED OR WITHHELD BASED UPON THE FALSE RESULTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER. CLINICAL CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. DXC600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1