FDA Adverse Event Malfunction Summary report: N

RENEGADE FIBER BRAIDED MICROCATHETER

MDR report key: 2080696 · Received May 6, 2011

Report

Report Number
2134265-2011-01562
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
January 12, 2011
Report Date
April 13, 2011
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
KRA
PMA / PMN Number
K020012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: RETURNED PRODUCT ANALYSIS REVEALED THE DEVICE WAS RETURNED WITHIN THE DISPENSER HOOP. SLIGHT RESISTANCE WAS NOTED WHEN ATTEMPTING TO REMOVE THE CATHETER. THE DISPENSER HOOP WAS FLUSHED WITH SALINE AND THE CATHETER WAS THEN REMOVED WITH EASE. VISUAL AND TACTILE TESTING WERE PERFORMED ON THE CATHETER. PEELING COATING WAS NOTED ON THE CATHETER SHAFT AT 74.3CM FROM THE PROXIMAL END. THE OUTER DIAMETER OF THE DEFECT WAS 0.0535. AN APPROPRIATE SIZED MANDREL WAS ADVANCED THROUGH THE HUB WITH NO RESTRICTIONS. ALL DIMENSIONS WHICH WERE MEASURED WERE FOUND TO BE WITHIN SPECIFICATION. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED USER ERROR AS CONTINUOUS FLUSH WAS NOT MAINTAINED AS INSTRUCTED IN THE DFU. (B)(4)

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2011. IT WAS REPORTED THAT DURING A RESERVOIR TREATMENT PROCEDURE, THE MICROCATHETER WOULD NOT INSERT INTO THE SHEATH. THE INTENDED TARGET LOCATION WAS THE LIVER. A 5FR NON BSC INTRODUCER SHEATH WAS PLACED. THE RENEGADE MICROCATHETER WAS ATTEMPTED TO BE INSERTED INTO THE SHEATH, BUT WAS UNSUCCESSFUL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME RENEGADE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD. HOWEVER, DEVICE ANALYSIS REVEALED THE COATING OF THE MICROCATHETER HAD PEELED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RENEGADE FIBER BRAIDED MICROCATHETER CATHETER, CONTINUOUS FLUSH KRA BOSTON SCIENTIFIC - CORK M001182520 13848718

Patients

Seq Age Sex Outcome Treatment
1 5FR MEDIKIT INTRODUCER SHEATH| 0.014" TRANSEND GUIDEWIRE