TALENT THORACIC STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-00867
- Event Type
- Injury
- Date Received
- April 22, 2011
- Date of Event
- March 24, 2011
- Report Date
- March 24, 2011
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070007
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). RESULTS: (ENDOLEAK). RESULTS & CONCLUSION: (CARDIAC DETERIORATION), (STENT GRAFT WAS DEPLOYED OUTSIDE THE PT AND RELOADED INTO THE DELIVERY SYSTEM).
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A FUSIFORM THORACIC AORTIC ANEURYSM AT ZONE 4 APPROX ONE MONTH AGO. THE VESSELS WERE MODERATELY CALCIFIED AND MODERATELY TORTUOUS. THE PROXIMAL NECK DIAMETER IS 42 MM, THE DISTAL NECK DIAMETER IS 36MM. IT WAS NOTED THAT DURING PROCEDURE, THE PT'S BLOOD PRESSURE COULD NOT BE DECREASED AS THERE WAS DETERIORATION OF THE PT'S CARDIAC FUNCTION. IT WAS REPORTED THAT THE TW4036 WAS DEPLOYED OUTSIDE THE PT, AND RELOADED REVERSED INTO THE DELIVERY SYSTEM. THE STENT GRAFT WAS THEN IMPLANTED IN THE DISTAL PORTION OF THE ANEURYSM. THE SAME THING WAS DONE WITH A TAXF4040 AND IT WAS IMPLANTED DISTAL TO THE TW4036. AT THIS POSITION, THE DELIVERY SYSTEM WAS ALONG THE INNER CURVE OF AORTA. DURING THE EXPANSION OF TAXF4040, IT STARTED TO MISALIGN (REF MFR # 2953200-2011-00868). THEN A TAXF4040 WAS IMPLANTED TO THE TARGET THE AREA OF THE MISALIGNMENT. THERE WAS A TYPE III ENDOLEAK COMING FROM A GAP BETWEEN THE MISALIGNED STENT GRAFT AND THE TW4036. ANOTHER MFR'S STENT GRAFT WAS IMPLANTED AT THE PROXIMAL PORTION OF THE THORACIC AORTA. A TAXF4444 WAS IMPLANTED INTO THE PROXIMAL END OF THE OTHER MFR'S STENT GRAFT. A TAXF4242 WHICH WAS ALSO DEPLOYED OUTSIDE THE PT REVERSED AND RELOADED INTO THE DELIVERY SYSTEM WAS IMPLANTED TO COVER THE TYPE III ENDOLEAK AND FOR APPOSING THE MISALIGNED STENT GRAFT TO THE GRAFT WALL OF TW4036. THEN, TAXW4040 WAS IMPLANTED FOR COVERING THE JUNCTION OF THE OTHER MFR'S STENT GRAFT AND THE TW4036. THE TYPE III ENDOLEAK WAS CONFIRMED TO HAVE RESOLVED AND THE PROCEDURE WAS CONCLUDED. NO ADD'L CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT THORACIC STENT GRAFT SYSTEM | MIH | MEDTRONIC IRELAND | NA | V00542666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention |