ANEURX AAADVANTAGE STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-00869
- Event Type
- Injury
- Date Received
- April 22, 2011
- Date of Event
- March 24, 2011
- Report Date
- March 24, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL RESULTS AND CONCLUSION: (MIGRATION, ENDOLEAK). (DISEASE PROGRESSION OF THE AORTIC NECK).
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 4 YRS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNK. ONE MONTH AGO THE PT PRESENTED WITH SEVERE BACK PAIN. IT WAS REPORTED THAT A CT SCAN WAS PERFORMED AND DEMONSTRATED THAT THE STENT GRAF MIGRATED INTO THE ANEURYSM SAC AND THERE WAS A TYPE I ENDOLEAK PRESENT. THE ANEURYSM DIAMETER IS 5.5 CM. THE VESSELS HAD MILD CALCIFICATION, THE AORTIC NECK WAS SEVERELY TORTUOUS AND THERE WAS DISEASE PROGRESSION. THE AORTIC NECK MEASURED 19 MM AT THE RENAL ARTERIES AND 27 MM IN DIAMETER 2 CM BELOW THE RENAL ARTERIES. A 25X16 MM ENDURANT STENT GRAFT WAS IMPLANTED FOUR DAYS LATER AND SUCCESSFULLY RESOLVED THE ENDOLEAK AND MIGRATION. NO ADD'L CLINICAL SEQUELAE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA | NA | 1000384 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |