FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM

MDR report key: 2080593 · Received April 22, 2011

Report

Report Number
2953200-2011-00869
Event Type
Injury
Date Received
April 22, 2011
Date of Event
March 24, 2011
Report Date
March 24, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS AND CONCLUSION: (MIGRATION, ENDOLEAK). (DISEASE PROGRESSION OF THE AORTIC NECK).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 4 YRS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNK. ONE MONTH AGO THE PT PRESENTED WITH SEVERE BACK PAIN. IT WAS REPORTED THAT A CT SCAN WAS PERFORMED AND DEMONSTRATED THAT THE STENT GRAF MIGRATED INTO THE ANEURYSM SAC AND THERE WAS A TYPE I ENDOLEAK PRESENT. THE ANEURYSM DIAMETER IS 5.5 CM. THE VESSELS HAD MILD CALCIFICATION, THE AORTIC NECK WAS SEVERELY TORTUOUS AND THERE WAS DISEASE PROGRESSION. THE AORTIC NECK MEASURED 19 MM AT THE RENAL ARTERIES AND 27 MM IN DIAMETER 2 CM BELOW THE RENAL ARTERIES. A 25X16 MM ENDURANT STENT GRAFT WAS IMPLANTED FOUR DAYS LATER AND SUCCESSFULLY RESOLVED THE ENDOLEAK AND MIGRATION. NO ADD'L CLINICAL SEQUELAE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA NA 1000384

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention