FDA Adverse Event Injury Summary report: N

ENDURANT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2080584 · Received April 22, 2011

Report

Report Number
2953200-2011-00866
Event Type
Injury
Date Received
April 22, 2011
Date of Event
March 24, 2011
Report Date
March 24, 2011
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS & CONCLUSION: (SEVERELY TORTUOUS AND MILDLY ANGULATED ILIAC ARTERIES).

Description of Event or Problem · 1

AN ENDURANT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX ONE MONTH AGO. THE AORTIC NECK WAS STRAIGHT AND THE ILIAC ARTERIES WERE SEVERELY TORTUOUS WITH MILD CALCIFICATION. THERE WAS SLIGHT KINKING OF THE STENT GRAFT IN THE ILIAC ARTERIES BILATERALLY (REF MFR 2953200-2011-00865). THE PHYSICIAN ELECTED TO IMPLANT TWO PALMAZ STENTS WITHIN THE ILIAC LIMBS OF THE BIFURCATED STENT GRAFT EXTENSION AND THE CONTRALATERAL ILIAC STENT GRAFTS. NO ADD'L CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDURANT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC IRELAND NA V00807925

Patients

Seq Age Sex Outcome Treatment
1 87 YR Required Intervention