FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,10

MDR report key: 20805015 · Received November 28, 2024

Report

Report Number
0002023141-2024-03886
Event Type
Injury
Date Received
November 28, 2024
Date of Event
January 28, 2024
Report Date
April 15, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K072589
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) TSV4B10, (IMP, TSV,4.1MM, SBM,10) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS. THE IMPLANT'S COLLAR WAS FRACTURED, AND THERE WAS A FRACTURED SCREW STUCK INSIDE THE IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1240277. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1240277 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE IMPLANT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. H6: EVENT PROBLEM CODES ¿ PATIENT CODE: 1690 ABSCESS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED FRACTURE OF IMPLANT COLLAR. IMPLANT #44 REMOVED AND BONE FILLED. SYMPTOMS: ABSCESS AND PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2613366 IMP,TSV,4.1MM,SBM,10 DENTAL IMPLANT DZE ZIMMER DENTAL 1240277

Patients

Seq Age Sex Outcome Treatment
1 69 YR Female