FDA Adverse Event Other Summary report: N

DA VINCI SURGICAL SYSTEM

MDR report key: 2080493 · Received May 6, 2011

Report

Report Number
2955842-2011-00139
Event Type
Other
Date Received
May 6, 2011
Date of Event
April 11, 2011
Report Date
April 11, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K011002
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING (FSE) CONFIRMED A BROKEN CABLE ON THE PATIENT SIDE MANIPULATOR (PSM) 2 ARM. THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. THE PSM WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE CUSTOMER REPORTED PROBLEM AND FOUND THE AXIS 6 CABLE TO BE BROKEN. AS OF (B)(6) 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE STARTING A DA VINCI PROSTATECTOMY PROCEDURE, THE SITE FOUND A BROKEN CABLE ON A PATIENT SIDE MANIPULATOR (PSM) ARM. THE PATIENT WAS UNDER ANESTHESIA WHEN THE SURGICAL STAFF DECIDED TO ABORT THE PLANNED PROCEDURE. NO PATIENT HARM, INJURY OR ADVERSE OUTCOME WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS1200 A4.3P9

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SYSTEM INSTRUMENTS AND ACCESSORIES