DA VINCI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2011-00139
- Event Type
- Other
- Date Received
- May 6, 2011
- Date of Event
- April 11, 2011
- Report Date
- April 11, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K011002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING (FSE) CONFIRMED A BROKEN CABLE ON THE PATIENT SIDE MANIPULATOR (PSM) 2 ARM. THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. THE PSM WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE CUSTOMER REPORTED PROBLEM AND FOUND THE AXIS 6 CABLE TO BE BROKEN. AS OF (B)(6) 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT WHILE STARTING A DA VINCI PROSTATECTOMY PROCEDURE, THE SITE FOUND A BROKEN CABLE ON A PATIENT SIDE MANIPULATOR (PSM) ARM. THE PATIENT WAS UNDER ANESTHESIA WHEN THE SURGICAL STAFF DECIDED TO ABORT THE PLANNED PROCEDURE. NO PATIENT HARM, INJURY OR ADVERSE OUTCOME WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS1200 A4.3P9 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SYSTEM INSTRUMENTS AND ACCESSORIES |