FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2080439 · Received May 6, 2011

Report

Report Number
1423500-2011-05605
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT SURVEILLANCE CONTACTED THE HP'S WIFE ON (B)(4) 2011 REGARDING THE REPORTED SYSTEM ERROR. THE HP'S WIFE STATED THAT THE HP HAS BEEN DOING FINE SINCE THIS REPORT AND THERE HAD BEEN NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS. THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) WAS CONFIRMED AND IS DUE TO USE ERROR. PER THE COMPLAINT INFORMATION, THE CAUSE OF THE SE 2240 IS DUE TO THE PATIENT HAVING A CLAMP OPEN ON AN UNUSED SUPPLY LINE. THE LOT NUMBER IS UNKNOWN; THEREFORE, A BATCH REVIEW CANNOT BE PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION.A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DWELL 1. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HOME PATIENT (HP) AND ASSISTED TO CYCLE POWER THE HC MACHINE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE