FDA Adverse Event
Malfunction
Summary report: N
I-STAT EG7+ CARTRIDGE
MDR report key: 2080266
·
Received April 6, 2011
Report
- Report Number
- 2245578-2011-00044
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- February 16, 2011
- Report Date
- April 1, 2011
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- GKZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2011, AN ABBOTT POINT OF CARE DISTRIBUTOR ((B)(4)) BECAME AWARE OF A CUSTOMER WHO STATED THAT THEY WERE YIELDING A HIGH RATE OF CODE 20, FILLING ISSUES AND STAR-OUTS. THERE WAS NO INFO AT THE TIME OF THE YIELD RATES. ON (B)(6) 2011, ABBOTT POINT OF CARE BECAME AWARE OF THE YIELD ISSUE BY THE ABBOTT POINT OF CARE DISTRIBUTOR. THERE WERE NO YIELD RATES AVAILABLE AT THIS TIME. ON (B)(6) 2011, ABBOTT POINT OF CARE BECAME AWARE THAT THE YIELD RATE FOR CODE 20 WAS 72 OUT OF 200 CARTRIDGES. THERE WERE NO YIELD RATES AVAILABLE FOR STAR OUTS OR FILL ISSUES. THE INVESTIGATION IS PENDING. BASED ON THE INFO AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THE EVENT.