FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 2080220 · Received April 7, 2011

Report

Report Number
2242352-2011-00243
Event Type
Malfunction
Date Received
April 7, 2011
Date of Event
March 7, 2011
Report Date
March 9, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY WAS INSIDE THE DELIVERY TUBE AND THE SEAL WAS EXTENDED OUTSIDE THE TUBE. THE SEAL WAS INTACT. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THE DEVICE WAS VERY BLOODY. TO FURTHER INVESTIGATE, THE SEAL WAS PULLED OUT OF THE DELIVERY TUBE. NO ISSUES WERE OBSERVED DURING THIS FUNCTIONAL TEST. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL DID NOT DEPLOY PROPERLY" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, TWO HEARTSTRING III SEALS DID NOT DEPLOY PROPERLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS III PROXIMAL SEAL SYSTEM 3.8MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC. HSK-3038 25020282

Patients

Seq Age Sex Outcome Treatment
1 NA