FDA Adverse Event
Malfunction
Summary report: N
CENTRA BED
MDR report key: 2080189
·
Received April 6, 2011
Report
- Report Number
- 1824206-2011-02037
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 14, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT FOUND THE CAUSE TO BE THE BRAKE CABLE BEING PINCHED BETWEEN THE STIFFENER PLATE AND THE BEARINGS. THE ACCOUNT INSTALLED NEW STYLE BEARINGS AND ADJUSTED THE BRAKE AND STEER CASTERS TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THAT THE BRAKES ARE NOT HOLDING. NO ALLEGED INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTRA BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 850 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |