FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2080155 · Received April 6, 2011

Report

Report Number
2531779-2011-02321
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 27, 2011
Report Date
March 10, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. INVESTIGATION CONFIRMED THAT THE AUDIO BOLUS BUTTON COVER IS MISSING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE TICKING AND INTERMITTENTLY UNRESPONSIVE. A FAMILY MEMBER REPORTED THAT THE CONTRAST AND DOWN ARROW KEYPAD BUTTONS HAVE BEEN STICKING AND REQUIRE MULTIPLE HARD PRESSES TO OBTAIN A RESPONSE FOR THE PAST 6 WEEKS. SHE ALSO REPORTED THAT THE AUDIO BOLUS BUTTON COVER FELL OFF TODAY. THE FAMILY MEMBER DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT THE PT WEARS THE PUMP IN HIS POCKET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 15 YR