Malfunction
Summary report: N
BV LIBRA
MDR report key: 2080126
·
Received April 7, 2011
Report
- Report Number
- 3003768277-2011-00329
- Event Type
- Malfunction
- Date Received
- April 7, 2011
- Report Date
- March 9, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- IZL
- PMA / PMN Number
- K010762
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE FIELD SERVICE ENGINEER REPLACED THE HIGH TENSION TANK. THIS SOLVED THE PROBLEM.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT HIGH TENSION (HT) TANK NECESSARY FOR THE GENERATION OF IMAGES USING X-RAY BECAME DEFECTIVE AND NEEDED REPLACEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BV LIBRA | IZL | PHILIPS HEALTHCARE | 718022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |