ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2080074
·
Received April 6, 2011
Report
- Report Number
- 2531779-2011-02301
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Report Date
- March 9, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. DURING INVESTIGATION, THE DOWN ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE UP AND DOWN ARROWS ARE RESPONDING INTERMITTENTLY. IT WAS REPORTED THAT THERE IS NO WATER EXPOSURE TO THE PUMP AND THERE IS NO DAMAGE TO THE KEYPAD.