FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME
Report
- Report Number
- 1037905-2011-00210
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- February 8, 2011
- Report Date
- March 7, 2011
- Manufacturer
- WILSON-COOK MEDICAL, INC.
- Product Code
- KNS
- PMA / PMN Number
- K052051
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- UNKNOWN
Narratives
THE INFORMATION IN THIS REPORT WAS PROVIDED TO US BY THE COOK FACILITY IN (B)(4) ON BEHALF OF A DISTRIBUTOR IN (B)(6). (B)(4). EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. A SAMPLE TEST FROM THIS LOT COULD NOT BE CONDUCTED BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN INVENTORY. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRE-CURVED AND IS PROVIDED WITH A PRE-CURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRE-CURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRE-CURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRE-CURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: NO CORRECTIVE ACTION WARRANTED AT THIS TIME BECAUSE THIS REPORT COULD NOT BE VERIFIED. THIS TYPE OF OCCURRENCE HAS BEEN BROUGHT TO THE ATTENTION OF THE APPROPRIATE INTERNAL PERSONNEL IN AN EFFORT TO HEIGHTEN AWARENESS. NO FURTHER ACTION WAS TAKEN AT THIS TIME. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME. THE SPHINCTEROTOME WAS ADVANCED INTO POSITION AND THE CUTTING WIRE ORIENTATION WAS REPORTED TO BE INCORRECT, CANNULATING TO THE RIGHT BETWEEN 2 AND 3:00. ANOTHER SPHINCTEROTOME WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME | KNS, ENDOSCOPIC ELECTROSURGICAL UNIT AND ACCESSORIES | KNS | WILSON-COOK MEDICAL, INC. | W2950556 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACTIVE CORD (UNKNOWN TYPE)| ELECTROSURGICAL UNIT (UNKNOWN TYPE)| ENDOSCOPE (UNKNOWN TYPE) |