FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2080039 · Received April 6, 2011

Report

Report Number
3015876-2011-00306
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 10, 2011
Report Date
March 10, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REPORTED POWER PCB ASSEMBLY AND DETERMINED THE CAUSE OF THE REPORTED FAILURE TO BE A SHORTED INTEGRATED CIRCUIT CHIP, DESIGNATOR U5.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON WITH AC OR DC POWER. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA