LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 2080025
·
Received April 6, 2011
Report
- Report Number
- 3015876-2011-00303
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 7, 2011
- Report Date
- March 7, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE SYSTEM CONTROLLER AND USER INTERFACE PCB ASSEMBLIES AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE POWERED ON, BUT THE DISPLAY WAS WHITE WITH PURPLE AND RED VERTICAL BARS. NO OTHER FUNCTIONS WERE AVAILABLE WHEN THE DEVICE WAS IN THIS STATE. THERE WAS NO REPORT OF PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |