FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 2080025 · Received April 6, 2011

Report

Report Number
3015876-2011-00303
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 7, 2011
Report Date
March 7, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE SYSTEM CONTROLLER AND USER INTERFACE PCB ASSEMBLIES AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE POWERED ON, BUT THE DISPLAY WAS WHITE WITH PURPLE AND RED VERTICAL BARS. NO OTHER FUNCTIONS WERE AVAILABLE WHEN THE DEVICE WAS IN THIS STATE. THERE WAS NO REPORT OF PT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA