ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-02334
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Report Date
- March 8, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT # B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
ON (B)(6), 2011, THE PT CONTACTED ANIMAS ALLEGING THAT SHE HAD CARTRIDGES THAT LEAKED. THE PT DENIED THAT HER BLOOD GLUCOSE (BG) WAS AFFECTED BY THE ALLEGED ISSUE. SHE INDICATED THAT SHE DID NOT SAVE THE CARTRIDGES. BASED ON THE INFO PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PT CLAIMED THAT SHE HAD CARTRIDGES THAT LEAKED. THE PT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | IR 1200 / 1250 / 2020/ OTP | B201581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |