OSSEOTITE XP IMPLANT 4/5 X 13MM
Report
- Report Number
- 0001038806-2024-02763
- Event Type
- Injury
- Date Received
- November 27, 2024
- Report Date
- April 9, 2025
- Manufacturer
- BIOMET 3I
- Product Code
- DZE
- PMA / PMN Number
- K063286
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM D4: ADDITIONAL DEVICE INFORMATION D9: DEVICE AVAILABILITY G3: DATE RECEIVED BY MANUFACTURER G6: TYPE OF REPORT H1: TYPE OF REPORTABLE EVENT H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER H4: DEVICE MANUFACTURER DATE H6: ADVERSE EVENT PROBLEM H10: ADDITIONAL NARRATIVE. ZIMVIE RECEIVED ONE (1) OS4513, (OSSEOTITE XP® IMPLANT 4/5 X 13MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS. THE IMPLANT WAS OBSERVED FRACTURED AT THE COLLAR REGION. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. THE DHR WAS REQUESTED; HOWEVER, AN ELECTRONIC COPY IS NOT AVAILABLE FOR REVIEW. THE INVESTIGATION WILL BE REOPENED AND UPDATED UPON ANY NEW OR ADDITIONAL INFORMATION PROVIDED TO THIS COMPLAINT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT IS WITHIN SPECIFICATIONS. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 182190 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE IMPLANT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD ¿ PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
ZIMVIE COMPLAINT (B)(4). B3: DATE OF EVENT IS NOT PROVIDED / UNKNOWN. E1: INITIAL REPORTER¿S EMAIL ADDRESS IS NOT PROVIDED / UNKNOWN.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #36 FRACTURED AND WAS REMOVED. THE IMPLANT FAILED DUE TO THE FRACTURE AT THE HEAD. IT WAS NOT A BONE OR INFECTION ISSUE. SITE GRAFTED: ALLOGRAFT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2292757 | OSSEOTITE XP IMPLANT 4/5 X 13MM | DENTAL IMPLANT | DZE | BIOMET 3I | 182190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Male | Required Intervention |