PUMP MMT-1782K 670G V4.11 MM
Report
- Report Number
- 2032227-2024-276266
- Event Type
- Malfunction
- Date Received
- November 27, 2024
- Date of Event
- November 3, 2024
- Report Date
- January 31, 2025
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000264727
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- 003
Narratives
THE PUMP PASSED THE SELF TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM WAS FOUND ON: 11/02/2024 16:07:22.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING BOLUS DELIVERY. 11/03/2024 15:41:04.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING BOLUS DELIVERY. 11/03/2024 15:42:58.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING BOLUS DELIVERY. 11/03/2024 15:44:56.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING PRIME. 11/03/2024 15:55:48.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING BOLUS DELIVERY. 11/03/2024 18:01:44.000 ALARMALERTNOTIFICATION FAULTNUMBER = NO DELIVERY (7) DURING BOLUS DELIVERY. THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO BOLUS DELIVERY ANOMALY OR HISTORY ANOMALY NOTED. NO UNEXPECTED NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM NOTED. PLEASE SEE BELOW FOR PUMP ERROR(S)/ALARM(S) NOTED 2 DAYS PRIOR TO THE EVENT DATE 03-NOV-2024 IN THE FORMATTED HISTORY FILE. INSERT BATTERY ALARM WAS FOUND ON: 11/03/2024 18:29:54.000 POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. DURING VISUAL INSPECTION, CORROSION WAS FOUND ON THE BATTERY TUBE ASSEMBLY NOTED. THE PUMP WAS CUT OPEN AND FOUND CORROSION ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A SCRATCHED CASE AND A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT. THE PUMP PASSED ALL THE REQUIRED TESTING. CUSTOMER ALLEGED FOR NO DELIVERY ALARM/INSULIN FLOW BLOCKED ALARM WAS NOT CONFIRMED. HOWEVER, DURING VISUAL INSPECTION CORROSION WAS FOUND ON THE BATTERY TUBE ASSEMBLY, PCBA 1, PCBA 2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
SELECT PATIENT INFORMATION CANNOT BE PROVIDED, DUE TO REGIONAL PRIVACY REGULATIONS. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW, ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK, BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED, TO MEDTRONIC MINIMED. THAT THE CUSTOMER, EXPERIENCED AN INSULIN FLOW BLOCKED ALARM. THE CUSTOMER REPORTED, NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1782K. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1782K WAS REQUESTED. AND CUSTOMER RESPONSE WAS, THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2452747 | PUMP MMT-1782K 670G V4.11 MM | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1782K | HG5L8PR | 000000763000264727 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Unknown |