FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ PMIC/ID-109

MDR report key: 20792024 · Received November 27, 2024

Report

Report Number
1119779-2024-00933
Event Type
Malfunction
Date Received
November 27, 2024
Date of Event
October 30, 2024
Report Date
January 31, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904486090
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1. INITIAL REPORTER PHONE #: (B)(6). THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K021954, K022172, K023273, K023301, K023568, K024152, K030677, K031306, K031679, K031679, K032131, K033784, K033907, K040006, K040106, K040716, K050089, K051689, K053241, K060214, K060217, K060218, K060218, K060493, K082538, K082852, AND K131331. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR MISIDENTIFICATION OF STAPHYLOCOCCUS AUREUS AS STAPHYLOCOCCUS EPIDERMIDIS WHEN USING PHOENIX PANEL PMIC/ID-109 (CATALOG NUMBER 448609) BATCH NUMBER 4123688. THE CUSTOMER DID NOT RETURN PANELS OR ISOLATES BUT PROVIDED PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB REPORTS SHOW S. AUREUS AND S. EPIDERMIDIS IDENTIFICATIONS WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, RETENTION PANELS OF THE COMPLAINT BATCH WERE TESTED USING QC ISOLATES S. AUREUS A29213 AND S. AUREUS A25923 ON A PHOENIX M50 MACHINE AND EVALUATED FOR IDENTIFICATION RESULTS. IN ADDITION, ONE CONTROL PANEL FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE TESTED USING QC ISOLATE S. AUREUS A29213 ON A PHOENIX M50 MACHINE AND EVALUATED FOR IDENTIFICATION RESULTS. THE PANELS TESTED IDENTIFIED THEIR INOCULATED ISOLATES AS S. AUREUS. THIS COMPLAINT IS NOT CONFIRMED. THE BATCH HISTORY RECORD WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-109 A PATIENT ISOLATE S. AUREUS WAS MISIDENTIFIED AS S. EPIDERMIDIS, THE USER NOTED THAT A CORRECTED REPORTED WAS MADE TO THE PROVIDER. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-109 A PATIENT ISOLATE S. AUREUS WAS MISIDENTIFIED AS S. EPIDERMIDIS, THE USER NOTED THAT A CORRECTED REPORTED WAS MADE TO THE PROVIDER. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
826446 BD PHOENIX¿ PMIC/ID-109 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) 4123688 30382904486090

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown