BD INTIMA-II 20GAX1.16IN PRN SLM
Report
- Report Number
- 3002601200-2024-00658
- Event Type
- Malfunction
- Date Received
- November 26, 2024
- Date of Event
- October 29, 2024
- Report Date
- December 9, 2024
- Manufacturer
- BD SUZHOU (MDS)
- Product Code
- FOZ
- UDI-DI
- 00382903830077
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
1. DHR/BHR REVIEW LOT#4081482. 1-THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 4 IN APRIL 2024, AND PACKAGED AT R240 PACKAGE LINE IN APRIL 2024. WORK ORDER QUANTITY WAS 99,000 EA. 2- REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MET THE PRODUCT SPECIFICATIONS. 3- REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLE AND PHOTO HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR RELATED TESTING: 800MM SIMULATED CLINICAL LEAKAGE TEST. THE TEST IS PASSED, AND NO LEAKAGE IS FOUND AT THE SEPTUM. PLEASE SEE THE ATTACHED TEST REPORT. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): SINCE NO ABNORMALITIES ARE FOUND IN THE PROCESS AND RETAINED SAMPLE, NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS, AND NO DEFECT STATUS OF THE COMPLAINED SAMPLE CAN BE IDENTIFIED, THE ROOT CAUSE OF THE LEAKAGE AT THE SEPTUM CANNOT BE DETERMINED.
IT WAS REPORTED THAT BD INTIMA-II 20GAX1.16IN PRN SLM LEAKAGE AT SEPTUM DURING USE OF THE PRODUCT, THERE WAS BLEEDING AND OOZING FROM THE ISOLATION PLUG; THE AFFECTED QUANTITY WAS 1 PLUG, THE SAMPLE CANNOT BE RETURNED, AND A PHOTO IS PROVIDED; A GREEN CLAIM IS REQUIRED, AND A COMPLAINT RESPONSE LETTER AND COMPLAINT ACCEPTANCE LETTER ARE NOT REQUIRED.
ADDITIONAL INFORMATION PROVIDED 11/22/2024 1. PLEASE PROVIDE PICTURE / VIDEO / PHYSICAL SAMPLE IF AVAILABLE. A: NO SAMPLES RETURNED. 2. THE BATCH NUMBER PROVIDED FOR THE RESPECTIVE MATERIAL NUMBER DO NOT MATCH IN SAP, PLEASE PROVIDE A VALID MATERIAL AND BATCH NUMBER. A: ITEM NUMBER: 383012, BATCH NUMBER: 4081482. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 878604 | BD INTIMA-II 20GAX1.16IN PRN SLM | INTRAVASCULAR CATHETER | FOZ | BD SUZHOU (MDS) | 4081482 | 00382903830077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |