TRUSTEEL
Report
- Report Number
- 3003442380-2024-32790
- Event Type
- Malfunction
- Date Received
- November 26, 2024
- Date of Event
- October 24, 2024
- Report Date
- June 26, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018457
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR(B)(4) DEVICE 2 OF 2
SUPPLEMENTAL REPORT 01 - MDR (B)(4). ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY THE INFORMATION IN THIS COMPLAINT 2051527 ACCORDING TO REFERENCE RESULTS COMPLAINT IS (B)(4) HAS BEEN EVALUATED. THE BATCH 6005559 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6005559 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE CODE TUBING DETACHED FROM TUBING-TUBING CONNECTOR. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL STATIC TEST 1 ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE PACKING LOT 6005559 WAS MANUFACTURED ACCORDING TO THE WI VERSION 96 MANUFACTURED IN THE MULTIVAC 10, ON 16/FEB/2024, WITH A TOTAL OF (B)(4) UNITS. THE GLUING TUBE LOT 4B01930 WAS MANUFACTURED ACCORDING TO THE WI VERSION 60 MANUFACTURED IN THE LINE SC08, ON 13/FEB/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 27/MAY/2025 AGAINST MALFUNCTION CODE TUBING DETACHED FROM TUBING-TUBING CONNECTOR. AND LOT 6005559 AND 1 COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, HARM REPORTED, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4) EVENT OCCURRED IN AUSTRALIA IT WAS REPORTED THAT THE PATIENT FACED TWO INFUSION SETS TUBING DETACHED EVENTS ON (B)(6)2024 AND (B)(6)2024 FROM CONNECTOR. INFUSION SETS WERE USED FOR 1.5 DAYS. BLOOD GLUCOSE WAS 14 MMOL/L AT THE TIME OF THE EVENT. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2357218 | TRUSTEEL | UNO CONTACT DETACH G29 80/6TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002834 | 6005559 | 05705244018457 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |