FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ PMIC/ID-106

MDR report key: 20787709 · Received November 26, 2024

Report

Report Number
1119779-2024-00932
Event Type
Malfunction
Date Received
November 26, 2024
Date of Event
November 8, 2024
Report Date
December 11, 2024
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904486069
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UNKNOWN LOT NUMBER: D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.4. DEVICE MANUFACTURE DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K021954, K022172, K023273, K023301, K024152, K030677, K031306, K031679, K031679, K032131, K033784, K033907, K040006, K040106, K040716, K050555, K051689, K053241, K060214, K060217, K060218, K060218, K060493, K082538, K082852, K082913, AND K131331. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY THIS COMPLAINT IS FOR HIGH MIC VANCOMYCIN WHEN USING PHOENIX PANEL PMIC/ID-106 (CATALOG NUMBER 448606) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT RETURN PANELS, ISOLATES OR PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-106 AN UNSPECIFIED NUMBER OF PATIENT ISOLATES HAD A HIGH MIC FOR THE DRUG VANCOMYCIN. THE USER NOTED REPEAT TESTING WAS PERFORMED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ PMIC/ID-106 AN UNSPECIFIED NUMBER OF PATIENT ISOLATES HAD A HIGH MIC FOR THE DRUG VANCOMYCIN. THE USER NOTED REPEAT TESTING WAS PERFORMED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
844108 BD PHOENIX¿ PMIC/ID-106 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) UNKNOWN 30382904486069

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown