IMP,TSV,MCOL MG,3.7MM,10M
Report
- Report Number
- 0002023141-2024-03846
- Event Type
- Injury
- Date Received
- November 26, 2024
- Date of Event
- October 1, 2024
- Report Date
- November 26, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019720
- PMA / PMN Number
- K111889
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT: UNKNOWN / NOT PROVIDED D10: CONCOMITANT MEDICAL PRODUCT: TSVM4B10, IMP,TSV,MCOL MG,4.1MM,10M, LOT: 1233821. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
DOCTOR REPORTED THAT IMPLANTS AT TOOTH SITES 12, 15 WERE REMOVED DUE TO PERI-IMPLANTITIS. INFLAMMATION WAS REPORTED AS A RESULT OF THE EVENT. PATIENT HAS BEEN RESCHEDULED FOR TREATMENT AND NEW IMPLANT PLACEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 844065 | IMP,TSV,MCOL MG,3.7MM,10M | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1234903 | 00889024019720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male | DENTAL IMPLANT SEE H10 NARRATIVE. |