FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER

MDR report key: 20778976 · Received November 26, 2024

Report

Report Number
1024879-2024-01137
Event Type
Malfunction
Date Received
November 26, 2024
Date of Event
October 26, 2024
Report Date
October 31, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686501
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 368650 AND LOT/BATCH #: 4192533. BD RECEIVED 7 SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING AND THE INDICATED FAILURE MODE FOR HUB-COLLAR SEPARATION WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL PRODUCT SPECIFICATIONS WERE MET. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE HUB-COLLAR SEPARATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHILE PREPARING TO USE A BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER, THE HUB COLLAR ALONG WITH THE SAFETY SHIELD DETACHED FROM THE DEVICE. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENTS OR USERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2576598 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH PRE-ATTACHED HOLDER HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 4192533 50382903686501

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown