TRUE METRIX AIR
Report
- Report Number
- 1000113657-2024-00468
- Event Type
- Malfunction
- Date Received
- November 26, 2024
- Date of Event
- November 1, 2024
- Report Date
- November 25, 2024
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: (B)(6): USER'S TEST STRIP HAD POOR STORAGE NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. CUSTOMER ALSO STATED THAT THE METER GIVES 40-50 POINTS SPREAD. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 143, 166 AND 139 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80-100 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 11/15/2025 AND OPEN VIAL DATE WAS NOT DISCLOSED. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION (KITCHEN). THE CUSTOMER DID HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY: LOT ZC5790S. DURING THE CALL, A BACK-TO-BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 158 MG/DL DATE: (B)(6)2024 TIME:1107AM FASTING. RESULT 2: 158 MG/DL DATE: (B)(6)2024 TIME: 806AM FASTING AM. RESULT 3: 165 MG/DL DATE: (B)(6)2024 TIME:804AM FASTING AM. RESULT 4: 159 MG/DL DATE: (B)(6)2024 TIME:644PM FASTING. RESULT 5: 160 MG/DL DATE: (B)(6)2024 TIME:1215PM FASTING. RESULT 6: 143 MG/DL DATE: (B)(6)2024TIME: 910AM FASTING. RESULT 7: 166 MG/DL DATE: (B)(6)2024 TIME:1025AM FASTING. RESULT 8: 139 MG/DL DATE: (B)(6)2024TIME: 129PM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1708167 | TRUE METRIX AIR | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TRUE METRIX 50CTMG/DL | ZC5658S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |