FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM

MDR report key: 20772647 · Received November 25, 2024

Report

Report Number
3011649314-2024-00919
Event Type
Injury
Date Received
November 25, 2024
Date of Event
October 29, 2024
Report Date
October 13, 2025
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS; HOWEVER, THE DEVICE EVALUATION IS PENDING. AT THE COMPLETION OF THE INVESTIGATION A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE ANALYSIS CONCLUSION. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED; DOCUMENTATION INDICATES THE PRODUCT MET RELEASE CRITERIA. THE PROBABLE ROOT CAUSE OF THE EVENT HAS NOT YET BEEN IDENTIFIED. MANUFACTURER INTERNAL REFERENCE NUMBER: (B)(4). RELATED TO MANUFACTURER REPORT NUMBER: 3011649314-2024-00920 AND 3011649314-2024-00921.

Additional Manufacturer Narrative · 0

THE DEVICE WAS EVALUATED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS THE DHR WAS REVIEWED FOR LOT#6200495 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS THE STOCK PRODUCT FOR HAHN TAPERED IMPLANT LOT#6200495 IS NOT APPLICABLE FOR REVIEW SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE OR THE DEVICE REPORTED. THE IMPLANT WAS NOT VERIFIED TO BE A HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM (70-1154-IMP0006) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE THREADING OF THE IMPLANT. (SEE ATTACHED IMAGES). THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE AS A DIFFERENT PRODUCT WAS RETURNED. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." A ROOT CAUSE FOR FIT ISSUE COMPLAINT CANNOT BE EXPLICITLY DETERMINED. PROBABLE ROOT CAUSE IS IMPROPER SEATING OF THE IMPLANT DRIVER INTO THE INTERNAL HEX OF THE IMPLANT. IT IS UNCLEAR THE METHODS OF IMPLANT PLACEMENT USED DURING THE INITIAL PROCEDURE OR IF APPROPRIATE LOAD DISTRIBUTION WAS OBSERVED. IFU 570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE IMPLANT PLACEMENT SECTION: "STEP 2: INITIAL PLACEMENT - ENGAGE THE IMPLANT CONNECTION WITH THE APPROPRIATE DRIVER. WITH THE IMPLANT SECURELY ATTACHED TO THE DRIVER, SQUEEZE THE OPPOSING END OF THE HOLDER TO DISENGAGE THE IMPLANT FROM THE HOLDER. TRANSPORT THE IMPLANT TO THE PREPARED SITE AND INSERT INTO THE OSTEOTOMY. ROTATE CLOCKWISE WITH APPLIED PRESSURE TO ENGAGE THE SELF-TAPPING GROOVES. AVOID LATERAL FORCES, WHICH CAN AFFECT THE ANGULATION AND FINAL ALIGNMENT OF THE IMPLANT. STEP 3: ADVANCEMENT AND FINAL SEATING - CONTINUE THREADING THE IMPLANT INTO THE OSTEOTOMY SITE USING THE PREFERRED PLACEMENT METHOD. A MINIMUM TORQUE VALUE OF 35 NCM UPON FINAL SEATING INDICATES GOOD PRIMARY STABILITY." IFU 570 REV 4.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE METHODS OF IMPLANT PLACEMENT SECTION: "OPTION 1: HANDPIECE IMPLANT PLACEMENT - PLACE THE APPROPRIATE IMPLANT DRIVER INTO THE HANDPIECE. SEAT THE DRIVER INTO THE INTERNAL HEX CONNECTION OF THE IMPLANT AND PRESS FIRMLY TO FULLY ENGAGE THE CONNECTION. THREAD THE IMPLANT INTO THE OSTEOTOMY AT APPROXIMATELY 25 RPM UNTIL FULLY SEATED. OPTION 2: MANUAL IMPLANT PLACEMENT - ASSEMBLE THE ADJUSTABLE TORQUE WRENCH WITH THE SURGICAL ADAPTOR AND APPROPRIATE IMPLANT DRIVER. WITH THE IMPLANT THREADED SECURELY IN ITS SITE, SEAT THE DRIVER INTO THE INTERNAL HEX CONNECTION OF THE IMPLANT, AND PRESS FIRMLY TO FULLY ENGAGE THE CONNECTION. TURN THE WRENCH CLOCKWISE IN INCREMENTS OF APPROXIMATELY 90 DEGREES. AVOID LATERAL FORCES, WHICH CAN AFFECT FINAL ALIGNMENT OF THE IMPLANT." MANUFACTURER REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

SECTION G3: MANUFACTURER WAS INCORRECTLY CAPTURED IN PREVIOUS REPORT WAS CORRECTED IN THIS REPORT. MANUFACTURER REFERENCE: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT A HAHN TAPERED IMPLANT LACKED PRIMARY STABILITY DURING IMPLANT PLACEMENT ON TOOTH NUMBER 7. IT WAS REPORTED THAT THE IMPLANT HAD A FIT ISSUE WITH INSTRUMENT DEFORMATION THAT FRACTURED AT THE TIME OF PLACEMENT. PER THE REPORT THE HEALTHCARE PROVIDER DID NOT OBSERVE ANY PATIENT SYMPTOMS OR PERMANENT INJURY. PER THE REPORT THE DEVICE WAS REMOVED, AND IT WAS REPLACED WITH A PRODUCT SIMILAR TO THE COMPLAINT PRODUCT AND NO ADDITIONAL PROCEDURES WERE PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2221754 HAHN TAPERED IMPLANT Ø3.5 X 11.5 MM HAHN TAPERED IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0006 6200495

Patients

Seq Age Sex Outcome Treatment
1 49 YR Male Required Intervention