BD BBL¿ TAXO¿ A
Report
- Report Number
- 2647876-2024-00161
- Event Type
- Malfunction
- Date Received
- November 22, 2024
- Date of Event
- October 24, 2024
- Report Date
- December 30, 2024
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- JTO
- UDI-DI
- 30382902315521
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: A COMPLAINT INVESTIGATION DUE TO NO INHIBITION ZONE WITH BACITRACIN TAXO A CATALOG 231552 BATCH NO.: 4023697 WAS PERFORMED ON RETENTION SAMPLES. RETURNED GOODS WERE NOT RECEIVED FROM CUSTOMER. THE INVESTIGATION REQUIRED TO TEST PRODUCT FOR PERFORMANCE, VISUAL INSPECTION, AND BATCH RECORD REVIEW. RETENTION SAMPLES PERFORMED AS EXPECTED. NO DISCREPANCIES OBSERVED DURING THE VISUAL INSPECTION. BATCH RECORD REVIEW DID NOT SHOW ANY EVIDENCE OF CUSTOMER CLAIM. NO CORRECTIVE ACTION IS REQUIRED BASED ON INFORMATION EVALUATED AND SATISFACTORY RESULTS OBTAINED DURING THE QC TEST. COMPLAINT IS NOT CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING AN UNSPECIFIED NUMBER OF BD BBL¿ TAXO¿ A, DISCS, THERE IS NO ZONE OF INHIBITION WHEN TESTING WITH STREP PYOGENES SAMPLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED WHILE USING AN UNSPECIFIED NUMBER OF BD BBL¿ TAXO¿ A, DISCS, THERE IS NO ZONE OF INHIBITION WHEN TESTING WITH STREP PYOGENES SAMPLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2068554 | BD BBL¿ TAXO¿ A | DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION | JTO | BECTON DICKINSON CARIBE LTD. | 4023697 | 30382902315521 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |