Report Number
1811755-2011-01079
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
December 1, 2010
Report Date
December 1, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FIELD SERVICE TECH REMOVED AND REPLACED THE POWER CORD. THE UNIT PASSED ALL SAFETY TESTS AND WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

THE FIELD SERVICE TECH REPORTED THAT THE NEPTUNE ROVER'S POWER CORD HAD EXPOSED WIRES AND THE PLUG WAS MELTED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK