BD MAX¿ ENTERIC PARASITE PANEL
Report
- Report Number
- 3007420875-2024-00134
- Event Type
- Malfunction
- Date Received
- November 22, 2024
- Date of Event
- July 29, 2024
- Report Date
- August 14, 2025
- Manufacturer
- GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
- Product Code
- OOI
- UDI-DI
- 00382904429607
- PMA / PMN Number
- K220193
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D.2. ADDITIONAL MEDICAL DEVICE TYPE: PCH. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: THE COMPLAINT INVESTIGATION FOR DISCREPANT RESULTS WHEN USING THE BD MAX ENTERIC PARASITE PANEL KIT (REF. (B)(4)) LOTS 3200365, 3265834 AND 4017468 WAS PERFORMED BY THE REVIEW OF MANUFACTURING RECORDS, REVIEW OF CUSTOMER¿S DATA AND BY THE COMPLAINT¿S HISTORY REVIEW. CUSTOMER REPORTED UNUSUAL CURVES ASSOCIATED WITH POSITIVE GIARDIA (GLAMB) AND CRYPTOSPORIDIUM (CRYPTO) TARGETS THAT REPEATED NEGATIVE WHEN USING BD MAX¿ ENTERIC PARASITE PANEL KIT LOTS 3200365, 3265834 AND 4017468. REVIEW OF MANUFACTURING RECORDS OF THE BD MAX ENTERIC PARASITE PANEL INDICATED THAT THE LOTS 3200365, 3265834 AND 4017468 WERE MANUFACTURED ACCORDING TO SPECIFICATIONS AND MET PERFORMANCE REQUIREMENTS. CUSTOMER PROVIDED RUNS 2888, 2891, 2980 AND 3188 FROM INSTRUMENT CT1566 FOR INVESTIGATION. SAMPLES TESTED WITH ENTERIC PARASITE PANEL KIT LOT 3200365 (RUN 2888; POSITION B6 AND 2891; B11), KIT LOT 3265834 (RUN 2980; POSITIONS B2 AND B4) AND KIT LOT 4017468 (RUN 3188; POSITION B2) WERE TARGETED AS TO BE INVESTIGATED BY THE CUSTOMER. SAMPLES IN RUN 2888; POSITION B6, 2891; B11 AND 2980; B4 TESTED POSITIVE FOR THE GLAMB TARGET, SAMPLE IN RUN 3188; POSITION B2 TESTED POSITIVE FOR THE CRYPTO TARGET, AND SAMPLE IN RUN 2980; POSITION B2 TESTED POSITIVE FOR BOTH GLAMB AND CRYPTO TARGETS. ALL TESTED NEGATIVE UPON REPEAT. MANUAL PCR CURVE ADJUDICATION WAS CONDUCTED ACROSS THE POSITIVE SAMPLES TARGETED. ANALYSIS REVEALED STEP DISLOCATIONS IN THE RAW PCR SIGNAL IN EVERY CHANNEL, RESULTING IN POSITIVE RESULTS FOR FOUR OF THE SAMPLES (RUNS 2891; POSITION B11, 2888; B6, 2980; B2 AND 3188; B2). IT IS UNLIKELY THESE POSITIVE RESULTS WITH ATYPICAL CURVES ARE TRUE AMPLIFICATIONS. SAMPLE IN RUN 2980; POSITION B4 SHOWED CURVES THAT SUGGEST A TRUE POSITIVE RESULT. MANUAL CURVE ADJUDICATION HAS LIMITATIONS; VISUAL EXAMINATION OF PCR CURVES FOR ABERRANT CURVE GEOMETRY IS A CONSERVATIVE ASSESSMENT OF THE DATA. THE ISSUES REPORTED OCCURRED OVER A PERIOD OF SIX MONTHS, USING THREE DIFFERENT KIT LOTS, SUGGESTING THAT THE ISSUE IS NOT LINKED TO SPECIFIC REAGENTS. THERE IS NO INDICATION OF A REAGENT ISSUE BASED ON THE ANALYSIS OF THE COMPLAINTS RECEIVED FOR DISCREPANT RESULTS ON BD MAX ENTERIC PARASITE PANEL LOTS 3200365, 3265834 OR 4017468. THE ROOT CAUSE WAS NOT IDENTIFIED. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. BD DID NOT INITIATE A CORRECTIVE AND A PREVENTIVE ACTION PLAN SINCE NO NEW HAZARD OR TRENDS WAS IDENTIFIED. BD QUALITY WILL CONTINUE TO MONITOR FOR TRENDS.
THE FOLLOWING FIELDS HAVE BEEN UPDATED: B5. IT WAS REPORTED WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT. THIS SUPPLEMENTAL MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023¿2025, UNDER CAPA 11910483.
REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, THREE (3) FALSE POSITIVE PATIENT RESULTS WERE OBTAINED. REPEAT TESTING GAVE NEGATIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.
REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, THREE (3) FALSE POSITIVE PATIENT RESULTS WERE OBTAINED. REPEAT TESTING GAVE NEGATIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.
IT WAS REPORTED WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1852933 | BD MAX¿ ENTERIC PARASITE PANEL | GASTROINTESTINAL PATHOGEN PANEL MULTIPLEX NUCLEIC ACID-BASED ASSAY SYSTEM | OOI | GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS) | 3265834 | 00382904429607 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |