GLENOSPHERE ORIENTATION GUIDE
Report
- Report Number
- 1818910-2024-24336
- Event Type
- Malfunction
- Date Received
- November 21, 2024
- Date of Event
- November 13, 2024
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- PXH
- UDI-DI
- 10603295116592
- PMA / PMN Number
- K170748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT COMPLAINT #(B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 230795000, LOT: 2753167 AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 230795000, LOT: 2753167 AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED.
DEPUY ORTHOPAEDICS IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY ORTHOPAEDICS HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY ORTHOPAEDICS OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY ORTHOPAEDICS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED ON (B)(6) 2024 THAT METAL TAB ON GLENOSPHERE ORIENTATION GUIDE BROKE OFF WHILE TRYING TO POSITION THE GLENOSPHERE ECCENTRICITY. SURGEON WAS UNABLE TO RETRIEVE THE BROKEN PIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1708005 | GLENOSPHERE ORIENTATION GUIDE | EXTREMITY INSTRUMENTS : ALIGNMENT DEVICES | PXH | DEPUY ORTHOPAEDICS INC US | 2753167 | 10603295116592 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|